Flowserve Logix MD+ ValveSight User Manual

Page 210

Advertising
background image

ValveSight™ Diagnostics DTM Manual for Logix MD+ Positioner with HART®

FCD-

LGENSF0014-00

© Flowserve Corporation

210


ILO Time Out

Definition: During calibration the Inner Loop Offset (ILO) value did not

settle. This could result in less accurate positioning.

Possible Solutions: Repeat the stroke calibration to get a more

accurate ILO value. To proceed using the less accurate ILO value, this
error may be cleared by briefly pushing the QUICK-CAL/ACCEPT
button. Lowering the setting on the gain selection switch may help if the
actuator is unstable during the calibration.

Stroke Shift

Definition: The 0% and 100% valve positions have both shifted in the

same direction since the last stroke calibration. This may be related to
a bent or adjusted feedback linkage, loose positioner mounting, or an
over rotated feedback potentiometer.

Possible Solutions: Ensure the feedback linkage is not bent and the

positioner is mounted securely. If the feedback potentiometer is over-
rotated, repeat the stroke calibration until the Stroke Shift error is no
longer present. This notification can be cleared by briefly pressing the
QUICK-CAL/ACCEPT button.

Stroke Span Increase

Definition: The 0% and 100% valve positions are farther apart

compared to the last stroke calibration. This could indicate seat wear.

Possible Solutions: Inspect valve or schedule valve for

inspection. This notification can be cleared by briefly pressing the
QUICK-CAL/ACCEPT button

Stroke Span Decrease

Definition: The 0% and 100% valve positions are closer together

compared to the last stroke calibration. This could indicate debris or
build up at valve seat.

Possible Solutions: Inspect valve or schedule valve for

inspection. This notification can be cleared by briefly pressing the
QUICK-CAL/ACCEPT button.


Command Input Calibration Flags

Command Range Small

Definition: During a Command Loop Calibration, the difference

between the signal at 0% and the signal at 100% was too small. The
system is designed to accept a difference greater than 5 mA.

Possible Solutions: Recalibrate making sure to use a larger

difference between command signal limits. The difference must exceed
5 mA.

Advertising
This manual is related to the following products: