TOHO ELECTRONICS TTM200 User Manual

Page 159

Advertising
background image

48-7009-E

6-11

■ PV filter setting

SEt02

INP2

Setting content

Initial value

6

PdF2

0.0 to 99.9 (seconds)

0.0


The PV filter setting is a function that demonstrates the CR filter effect on software through performing a first-order lag
calculation for PV of the remote SV input.

The filter effect is set using the time constant [t].

(Time constant is defined as the time required for PV to reach about 63 % when an input changes in a step configuration.)
* CR filter --- a filter of first-order lag

Note Application of PV filter

(1) Effect of noise is alleviated due to elimination of high-frequency noise when electric noise applies on inputs.
(2) Response can be delayed against abrupt change of an input.

100%

0%

Time

100%

0%

Time

100%

0%

Time

63%

Time constant (t)

■ Input signal




■ Input read in

When no digital PV filter exists

[Time constant (t) = 0]


When a digital PV filter exists

[Time constant (t) > 0]

■ Local/remote switchover

SEt02

INP2

Setting content

Initial value

7

LR

0

Local

0

1

Remote 1 (scaling at SLL , SLH)

2

Remote 2 (scaling at FSL 2, FSH 2)

Advertising