PR 5230 Instrument Manual

Commissioning

Sartorius

EN-103

5.3.6

Function INFO

When you press

the program releases and status messages are displayed. The

key also has other

functions; see Chapters 5.4.1.1 and 5.4.9.

Info

Show version

Show status

Show HW-slots

When you select [Show version], the installed program releases and the board number are displayed:

Info/

Version

Firmware

Rel. 03.00.00.00000 Firmware release and creation date

2011-10-10 11:11

Application

Rel. 01.20.05 Application release and creation date

2010-11-11 11:11

Bios

Rel 02.00.00.00000 BIOS release and creation date

2010-11-11 11:11

Boardnumber

327923612 Main board identification number

(different from the device serial number)

When you select [Show status], instrument status information is displayed:

Info/

Status

Free system RAM

5328 of 15204 kb

CAL-Switch

opened opened = CAL switch is open,

closed = CAL switch is closed.

Builtin RS232

no signal

When you select [Show HW-slots], the installed plug-in cards are displayed:

Info/

HW-Slots

Builtin

RS485 Firmly installed serial interface

Builtin

RS232 Firmly installed serial interface

Builtin

analog out Optional interface (Slot 1, B1*)

Builtin

digital i/o Firmly installed digital interface

PR1721/41

ProfiBus-DP Only field bus cards (Slot 2, B2*)

PR5230/xx

ADC For weighing electronics board (W1 or WE1)

When selecting ‚PR5230/xx’, factory serial number and production date are displayed:

Serial No 30252 00055

Prod.date: 2010-04-26 13:48:30

5.3.7

Setup Function (VNC)

Press

to configure the main operating parameters. The configuration depends on the application licenses

registered and the plug-in cards installed.
Calibration is in a simple dialogue. Compliance with the relevant (verification) standards must be checked by
the person commissioning the instrument or the verification officer. To protect the calibration data from
overwriting, close the corresponding CAL switch on the weighing electronics board. On legal-for trade
instruments, the CAL switch must be sealed in the closed (write-protected) position; see Chapter 5.1.1.