15 error messages, 1 error messages in measuring circuit, Error messages – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 214: Error messages in measuring circuit, Pr 5230 instrument manual

Advertising
background image

Error Messages

PR 5230 Instrument Manual

EN-214

Sartorius

15

Error Messages

15.1

Error Messages in Measuring Circuit

The weighing electronics can generate error messages, which are output on the weight display.

Indication VNC text

Error and possible cause

Remote display

E:Arith

Arith. error

Internal arithmetic overflow:
-

Faulty calibration values

Error 1

E:Overl

Overload

Input voltage is higher than Max + (x d):
-

faulty setting

-

to much weight on the scale

Error 2

E:NoSig

Ext. meas.device error

Measuring input open:
-

Measuring cable is interrupted (Cable break

detection)

Error 3

E:Under

Value exceeds display

OIML selected: Gross weight < -20d
-

The scale has been unloaded excessively

Error 4

E:Sense * No sense voltage

No sense voltage:
-

Load cells not connected.

-

Sense line or supply line is interrupted.

-

Wrong polarity or sense voltage is low

Error 6

E:Invers

Negative input

Negative input:
-

Wrong polarity of load cell signal.

-

Wrong polarity of supply voltage.

Error 7

E:HardE

No values from scale

Internal weighing point: Input signal is higher
than the permissible range of 36 mV.
Cannot read weight values from ADC
(analogue-digital-converter).
-

Error in weighing electronics board

-

Defective load cell

-

Cable break

Error 9