15 error messages, 1 error messages in measuring circuit, Error messages – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 214: Error messages in measuring circuit, Pr 5230 instrument manual
Error Messages
PR 5230 Instrument Manual
EN-214
Sartorius
15
Error Messages
15.1
Error Messages in Measuring Circuit
The weighing electronics can generate error messages, which are output on the weight display.
Indication VNC text
Error and possible cause
Remote display
E:Arith
Arith. error
Internal arithmetic overflow:
-
Faulty calibration values
Error 1
E:Overl
Overload
Input voltage is higher than Max + (x d):
-
faulty setting
-
to much weight on the scale
Error 2
E:NoSig
Ext. meas.device error
Measuring input open:
-
Measuring cable is interrupted (Cable break
detection)
Error 3
E:Under
Value exceeds display
OIML selected: Gross weight < -20d
-
The scale has been unloaded excessively
Error 4
E:Sense * No sense voltage
No sense voltage:
-
Load cells not connected.
-
Sense line or supply line is interrupted.
-
Wrong polarity or sense voltage is low
Error 6
E:Invers
Negative input
Negative input:
-
Wrong polarity of load cell signal.
-
Wrong polarity of supply voltage.
Error 7
E:HardE
No values from scale
Internal weighing point: Input signal is higher
than the permissible range of 36 mV.
Cannot read weight values from ADC
(analogue-digital-converter).
-
Error in weighing electronics board
-
Defective load cell
-
Cable break
Error 9