Procedure on the notebook/pc, Device presettings in the setup menu, Pr 5230 instrument manual extended functions – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 209
PR 5230 Instrument Manual
Extended Functions
Sartorius
EN-209
Procedure on the Notebook/PC
4.
Click the relevant file in the ‘Explorer’ with the mouse and drag it into window [Select
]
(or use ‘Copy and paste’).
5.
Select [Network].
6.
Fill in the device IP address.
7.
Click [Flash] to start the procedure.
As soon as software loading is completed the device will be re-started.
Now the next file (firmware) can be loaded as described in steps 2…7.
12.2.3
Updating the Software in a Network with a Fixed IP Address
Prerequisites:
-
Device and notebook/PC are connected to each other.
-
The automatic address assignment ‘DHCP’ is deactivated in the device and in the
notebook/PC, see Chapter 18.2/18.3.
-
The ‘FlashIt!32’ program (in a directory on the enclosed CD-ROM) is installed on the
notebook/PC.
-
The ‘FlashIt!32’ program is started.
Device presettings in the setup menu
1.
Select
-[Network parameter].
Setup/
Network parameter
HW address
00:90:6C:6B:6A:5E
Hostname
PR5230-6A6B5E
Use DHCP
IP Address
172.24.22.5 2.
Enter the corresponding IP address.
Subnetmask
255.255.0.0 3.
Enter the corresponding subnet mask.
Default gateway
0.0.0.0
Remote access
VNC-Client
255.255.255.255
4.
Leave with
.
The IP addresse is shown in the display of the device.