11 determination test value/display test value, 12 finishing/saving the calibration, Determination test value/display test value – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 121: Finishing/saving the calibration, 12, in, Pr 5230 instrument manual commissioning

Advertising
background image

PR 5230 Instrument Manual

Commissioning

Sartorius

EN-121

5.4.11

Determination Test Value/Display Test Value

Press [Calc Test] activates test value calculation. The maximum capacity (Max) is displayed with the
designation

without a weight unit.

The value determined during calibration after starting the test [CalcTest] is shown.

Weighingpoint/WP A/

Calibration

Calc Test

Depending on the settings under

-[Weighingpoint]-[Calib]-[Param]-[Test mode], either

-

the maximum capacity with [Absolute], or

-

the deviation from the test value with [Relative]

is shown when you press

to view the test data.

5.4.12

Finishing/Saving the Calibration

Finish the calibration with

.

You are prompted to confirm whether calibration should be closed without
determining the test value.

If not all data was determined when calibrating with [New] (e.g. dead load
not set/entered), this message is shown:
Press [Yes] to confirm and then press again; another prompt is displayed:

Press [Save] to save changes in calibration data. If you press [Undo], changes
are not saved and the display returns to the selection menu for the
weighing points.

The confirmation is displayed as follows:

Saving calibration

After quitting the menu, the following message is displayed:

Exit calibration

After finishing the calibration, set the CAL switch to the closed position; see also Chapter 5.1.1.

Advertising