2 updating a new software with ‚flashit, 1 updating in a network using dhcp service, Operations on the instrument – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 206: Updating a new software with ‚flashit, Updating in a network using dhcp service, Extended functions, Pr 5230 instrument manual
Extended Functions
PR 5230 Instrument Manual
EN-206
Sartorius
12.2
Updating a new Software with ‚FlashIt’
Note:
Always flash/load the BIOS into the device first, and then the firmware&application.
12.2.1
Updating in a Network Using DHCP Service
Prerequisites:
-
Device and notebook/PC are connected to a network.
-
The automatic address assignment ‘DHCP’ is activated in the device and in the
notebook/PC (default setting), see Chapter 18.2/18.3.
-
The ‘FlashIt!32’ program (in a directory on the enclosed CD-ROM) is installed on the
notebook/PC.
-
The ‘FlashIt!32’ program is started.
Operations on the instrument
Danger! High Voltage!
Working on the instrument while it is switched on may have life-threatening
consequences.
The operation may be performed only by trained and qualified persons who are familiar
with the implied hazards and avoid them, or take adequate measure to protect
themselves.
1
2
3
1.
Open the door of the instrument.
2.
Press keys (1) and (3) simultaneously and press key (2) three times.
On the display appears:
3.
Press the key (1) to confirm the loading of the software.