PR 5230 Instrument Manual

Installing the Instrument and Plug-in Cards

Sartorius

EN-47

4.2.9

Interface RS-485

The instrument is provided with a built-in RS-485/422 interface.
When mounting, the RS-485 interface must be configured by DIL switch S401 on the card.
Using RS-485 is compulsory with a multi-point connection (Tristate status). The RS-485 interface can be used
also for point-to-point connection.

Connection

5-pin plug-in terminal block

Number of channels

1 RS-485

Type

Full duplex

Transfer rate

300…19,200 bit/sec

Signals

TxA, RxA, TxB, RxB

Electrical isolation

Yes

Cable length

Max. 1000 m

Cable type

shielded twisted pair
(e.g. LifYCY 3x2x0.20),
1 pair of wires for GND

DIL switch S401

Block diagram RS-485

TxB

RxD/TxD-P

RxD/TxD-N

TxA

RxB

RxA

4

5

3

2

1

GND

S4

S1

S2

S6

S7

S3

1K6

1K6

120

120

S5

DC

DC

1 2 3 4 5 6 7 8

ON

DIP

S

Function

Settings for RS-485
4-wire


2-wire

1

Rx bus termination

OFF: not connected

ON: (RxA 120 ¥ RxB)

2

Rx pull-up resistor

OFF: not connected

ON: (RxB 1K6 ¥ +V)

3

Rx pull-down resistor

OFF: not connected

ON: (RxA 1K6 ¥ -V)

4

Tx/Rx bus termination

OFF: not connected

ON: (TxA 120 ¥ TxB)

5

Change-over 2- / 4-wire

OFF: 4-wire

ON: 2-wire

6

Rx enable

OFF: Rx disabled, if Tx enabled:

S7 must be ON.

ON: Rx always enabled: For

4-wire S7 must be OFF.

7

Tx enable/Rx disable (2-wire)

OFF: S6 must be ON.

ON: S6 must be OFF.

8

n. c.

no function