Adapting the analog output, Pr 5230 instrument manual commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 159
PR 5230 Instrument Manual
Commissioning
Sartorius
EN-159
5.10.1
Adapting the Analog Output
The output current can be adapted in small ranges. This is required, if small deviations from the nominal value
occur in a connected PLC.
Press
-[Show HW-slots] to open the menu:
Info/
HW-Slots
Builtin
RS485
Builtin
RS232
Builtin
analog out
Builtin
Digital i/o
PR1721/45
CC-Link
PR5230/xx
ADC
Press to select [analog out].
Info/
HW-Slots
Builtin
analog out
In use by PLC task
2
Analog output
0.000 mA
counts
12 cnt
Stop PLC Stop I/O Adjust
Reset
Press [Adjust]
Adapt analog output ?
Safety prompt: Reply [Yes] to start adapting.
Info/HW-Slots/
Adjust Analog Output
Output
4.000 mA
Measured
4.004 mA
Enter e.g. the value for 4 mA measured by the connected PLC under [Measured].
After pressing
, the 2nd value (20 mA) is displayed:
Info/HW-Slots/
Adjust Analog Output
Output
20.000 mA
Measured
20.010 mA
Enter e.g. the value for 20 mA measured by the connected PLC under [Measured].