6 word addresses, Word addresses, Pr 5230 instrument manual j-bus/modbus protocol – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 175

Advertising
background image

PR 5230 Instrument Manual

J-Bus/ModBus Protocol

Sartorius

EN-175

Example of an error message, which was generated by an invalid function number with ModBus-RTU.
Command

1

9

0

0

0

0

CRC

CRC

Reply

1

137

1

CRC

CRC

Example for ModBus-TCP:
Command

47

16

0

0

0

8

0

9

0

0

0

0

Reply

47

16

0

0

0

3

0

137

1

The individual bytes are shown.

6.6

Word Addresses

16 Gross weight, 1

st

byte (MSB)

Gross weight, 2

nd

byte

17 Gross weight, 3

rd

byte

Gross weight, 4

th

byte (LSB)

2

39

38

37

36

35

34

33

32

47

46

45

44

43

42

41

40

7

119 118 117 116 115 114 113 112 127 126 125 124 123 122 121 120

Address

Description

Read bits:

32

ADC error

33

Above Max (full scale deflection)

34

Above Max + n d

35

Below zero (minus sign)

36

Zero within ¼ d

37

Within zero setting range

38

The weight is stable

39

The weight is below zero or above Max

Write bits:

112

Set zero

113

Set tare

114

Reset tare

For further bits, see Chapter 10.

Advertising