Calibration of the Internal Weighing Point

Calibration using weights, mV/V or load cell data can be done using the VNC program. During calibration, the
instrument must be set to gross weight display (reset tare, if necessary).

For a ’legal-for-trade’ application, set the mode under

-[Calibration]-[Param] to [W

&M] before starting

calibration; see Chapter 5.4.13.1.
Select [New] to go to the maximum capacity [Max] (see Chapter 5.4.3); select the scale interval and determine
the dead load.
Now calibrate the maximum capacity by applying a calibration weight, in mV/V or with load cell data.
After determining the test value, the menu can be closed as described in Chapter 5.4.12, in order to save the
new settings.
Calibration data can be protected by the CAL switch (see Chapter 5.1.1), which must be sealed in the closed
(write-protected) position for ‘legal-for-trade’ applications.

5.4.1

Displaying Calibration Data

When the CAL switch is closed, the following message is displayed, only data
display possible with [Param]:

With [View], the calibration data can be displayed, but not changed.

WP A/

View Calibration

Max

3000 d

3000 kg

Number of scale intervals and max. capacity

Scale interval

3000 d

1 kg

Scale interval

Deadload at

0.00 kg

0.000000

Dead load in weight and mV/V

Max at

3000.00 kg

1.000000

Weight and mV/V for maximum capacity

Calibrated at

3000.00 kg

1.000000

Test load* and corresponding mV/V

Sensitivity

833.33

4.000000

Number of internal counts and voltage

per scale interval

Param

The calibration data and parameters (press [Param]) are displayed in the format entered/determined during
calibration.
* After input with mV/V, the maximum capacity and the mV/V value entered are displayed.

5.4.1.1

Increased Resolution (10-Fold)

In the

-[Calibration] menu, the weight is displayed with 10-fold resolution (also with the CAL switch

closed) when you press the key

, and marked as an invalid weight with

above the weight unit. The

display on the instrument remains unchanged. After 5 s, the VNC display returns to normal resolution or you
can press the key to return to normal display immediately.