1 status leds on fieldbus card, 2 status display, Display – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 85: Legend, 3 connection

Advertising
background image

PR 5230 Instrument Manual

Installing the Instrument and Plug-in Cards

Sartorius

EN-85

4.4.7.1

Status LEDs on Fieldbus Card

Watchdog LED

Meaning

Frequency

Color

Flashing

1 Hz green

Module is initialized and ready for operation.

Flashing

2 Hz green

Module is not initialized.

Flashing

1 Hz red

Check error in ASIC and FLASH ROM, module is defect.

4.4.7.2

Status Display

Requirements

In

-[Display items] the items are configured, see Chapter 5.6.7.

In

-[HW-Slots] PR 1721/45 is selected.

Display

1

2

3

4

LED ---

--- ---

---

1

2

3

4

Lit

- - -:

No power on the
module.

Lit

- - -:

No power on the
module.

Lit

- - -:

No power on the
module.

Lit

- - -:

No power on the
module.

Lit

grn:

Normal operation.

Lit

red:

CRC error,
Illegal station number or
illegal baudrate selected.

Lit

grn:

Data being transmitted.

Lit

grn:

Data being received.

Legend

---

off

red

red

grn

green

4.4.7.3

Connection

Caution!
The cable screen must be taken into the instrument through the cable gland and

connected to terminal 4 (see page 84).
Before, during and after installation, make sure that the sealing rings are seated

correctly.