4 determining the scale interval, Determining the scale interval, Commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 114: Pr 5230 instrument manual
Commissioning
PR 5230 Instrument Manual
EN-114
Sartorius
5.4.4
Determining the Scale Interval
The scale interval (d) is the difference between two successive display values.
With a legal-for-trade weigher, this value is called “verification scale interval“, which corresponds to the scale
interval:
d = e.
Example
Max = 3000 kg
Scale division = 1
Calculation (automatic):
d = Max / scale division
d = 3000 kg / 1 kg
d = 3000
Weighingpoint/WP A/
Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg The scale division 1, 2, 5 10, 20 or 50
can be set by pressing
.
The scale interval [d] is calculated from the scale division and the Max weight value.
The weight unit is taken from [Max] and cannot be changed here.
The number of digits behind the decimal point has been determined already when entering [Max] as well.
After
or confirmation of the change is displayed with:
Set Scale interval
This message displays, if the maximum capacity [Max] of the scale range
(weighing range) is not an integer multiple of the scale interval.