Asycom Protocol

PR 5230 Instrument Manual

EN-186

Sartorius

8

Asycom Protocol

This function enables the use of the following commands of the old EW-command set, e.g. PR 1612- command
set.
No license is required for using.

8.1

Commands of the Weight Function

Command

Reply

Description

WGA

QGAwwwwwwemz

Read gross weight

WNA

QNAwwwwwwemz

Read net weight

WTA

QTAwwwwwwemz

Read tare weight

wwwwwwe = Weight with sign and exponent
(e.g. '002340' = 002.34 g; 001005 = 100 kg)
e = Exponent
m = 1: no motion/Stand still
z = 1: tared with status request

Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E30000 BCC error
E50000 analog test active
E60000 weighing point unknown

WZA

Q

Set instrument to zero

Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E30000 BCC error
E60000 weighing point unknown
E70000 weighing point tared or no standstill

WSA

Q

Set tare

WSAwwwwwwe

Q

Fixtare

Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E30000 BCC error
E60000 weighing point unknown
E63000 Fixtare > MAX
E70000 weighing point tared or no standstill

WFA

Q

Reset tare

Possible error messages:
E100@0 general error, for @ see Chapter 8.3
E60000 weighing point unknown
E70000 weighing point tared or no standstill

ZS1Awwwwwwewwwwwwe QZS1A

Set Limit 1

ZS2Awwwwwwewwwwwwe QZS2A

Set Limit 2

ZS3Awwwwwwewwwwwwe QZS3A

Set Limit 3

Possible error messages:
E20000 error in telegram
E60000 weighing point unknown
E63000 Fixtare >MAX

I

QIwwwww ddssnnnnn d = Unit, variable (mg, g, kg, t, lb, oz)

ss = Scale interval (01, 02, 05, 10, 20, 50)
nnnnn = Span in mV/V (01234 acc. 0.1234 mV/V)