13 parameter input, Measuretime, Digital filter – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 122: Fcut, External supply, Test mode, W & m, Standstill time, Standstill range, Parameter input

Advertising
background image

Commissioning

PR 5230 Instrument Manual

EN-122

Sartorius

5.4.13

Parameter Input

Open the menu via

-[Weighingpoint]-[Calib]-[Param].

Weighingpoint/WP A/

Calibration

Measuretime

160 ms

Digital filter

bessel

Fcut

2.00 Hz This line is shown only, if the filter is on.

External supply

Above 8V

Test mode

absolute

W

& M

none

Standstill time

0.50 s

[Measuretime]
Select the measurement time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms.
In 'legal-for-trade' mode select

≤1 s.

[Digital filter]

The digital filter can be switched on only with the [measurement time] set to

≤160 ms.

Select the filter characteristic: [off], [Bessel], [aperiod.], [Butterw.], [Tscheby.].

[Fcut]
Enter the cut-off frequency for the low-pass filter (0.1…80 Hz); the setting is dependent on the measurement
time.

[External supply]

The external supply voltage is selected. Possible values: >8 V and

≤8 V.

[Test mode]
With [absolute], the test value is determined when the test is called. With [relative], the deviation from the
initially stored test value is displayed; see Chapter 5.4.10.

[W

& M]

Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia); see
Chapter 5.4.13.1.

[Standstill time]
Parameters [Standstill range] and [Standstill time] are required for determining the mechanical stability of the
scale. Input in seconds; permissible range: 0.01…2 sec. If 0 is set, stability is not checked. The stability time
must not be less than the measurement time.

[Standstill range]
The mechanical stability of the scale can be detected as long as any changes in the weight value are within
this range; permissible range 0.01…10.00d. In ‘legal-for-trade’ mode, select

≤1 d.