Scale data, Scale information, Escape command – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 180: 4 sma reply messages, Standard reply, Sma reply messages, Sma protocol, Pr 5230 instrument manual

Advertising
background image

SMA Protocol

PR 5230 Instrument Manual

EN-180

Sartorius

Scale Data

Scale Data – First Line

Command: <LF>A<CR>
Reply:

The scale sends the first line of its scale data.

<LF><SMA>:<yyyyyy><CR>

For details, see page 180.

Scale Data – Other Lines

Command: <LF>B<CR>
Reply:

The scale sends further lines of its scale data.

<LF><MFG>:<yyyyyy><CR>

For details, see page 180.

Scale Information

Scale Information - First Line

Command: <LF>I<CR>
Reply:

The scale sends the first line of its scale information.

<LF><SMA>:<yyyyyy><CR>

For details, see page 180.

Scale Information – Other Lines

Command: <LF>N<CR>
Reply:

The scale sends further lines of its scale information.

<LF><TYP>:<yyyyyy><CR>

For details, see page 180.

Escape Command

Command: <ESC>
Reply:

This is the only command which does not work according to the protocol principle. It does not
have a reply. The <ESC> character is detected at any time and cancels any current command.

7.4

SMA Reply Messages

In this section, the replies are described in detail. The data format of each reply has a fixed length. The
communication error is the only exception from this pre-definable format. Thus the controlling computer can
check each reply according to fixed rules, because each data field is in a fixed position.

Standard Reply
With most commands, the reply format is as described below:
Exceptions are the commands: 'D', 'A'/'B' and 'I'/'N'.
<LF> <s> <r> <n> <m> <f> <xxxxxx.xxx> <uuu> <CR>

Advertising