4 error messages of the calibration, Error messages of the calibration, Error messages – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 216: Pr 5230 instrument manual
Advertising
Error Messages
PR 5230 Instrument Manual
EN-216
Sartorius
15.4
Error messages of the Calibration
The maximum capacity is too high.
Subsequent changing of the maximum capacity is possible; if you
decrease the capacity, a message is displayed if the new maximum
capacity is lower than the test load ([Calibrated at]).
This message displays, if the selected resolution is to low, e.g. 5 kg.
The selected resolution is so high that less than 0.8 internal counts per
scale interval (d) or 0.5 µV/e for legal-for-trade acc. to OIML/NSC are
available.
The maximum capacity is not an integer multiple of the scale interval.
Weight units do not match, e.g. subsequent change of [Max] from kg to
lb
Advertising