8 pr 1721/46 profinet i/o interface, 1 status display, Display – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 86: Pr 1721/46 profinet i/o interface, Installing the instrument and plug-in cards, Pr 5230 instrument manual

Advertising
background image

Installing the Instrument and Plug-in Cards

PR 5230 Instrument Manual

EN-86

Sartorius

4.4.8

PR 1721/46 ProfiNet I/O Interface

The ProfiNet I/O interface PR 1721/46 is a plug-in card for mounting in slot 2, see Chapter 4.4.1. The card is
fitted with a standard RJ

-

45 socket for network connection and contains a powerful UDP/IP connecting

circuitry with transfer rates of 10 and 100 Mbits/s.

Internal connection

34- pin connector on flat cable for
slot 2

External connection

RJ-45 connecting socket

Transfer rate

10 Mbit/sec and 100 Mbit/sec
Autodetection (10/100, HalfDX/FullDX)

Connection mode

Network

Protocol

ProfiNet I/O

Configuration

XML file
‚GSDML-xxx-Sartorius-PR5230-xxx.xml’

Cable

Twisted pairs, screened
e.g. patch cable CAT5
Autolink (straight or crossover)

Cable impedance

150 ¥

Cable length to HUB

Max. 115 m

Certificate

ProfiBus Nutzerorganisation e.V.
for HMS Industrial Networks AB
Certificate No: Z10006
Report: PN005-1, 12.02.2007.

Potential isolation Yes
Dimens. (LxWxH) 87 x 55 x 15mm
Weight

125 g


Note:

The IP-Address and the Subnet-Mask are set at Fieldbus parameter (see also Chapter 5.6.5 and 9).
The XML file is stored on the CD (directory ‘Fieldbus’ of the according instrument) supplied with
the unit. The current file is also available for download via the Internet:
http://www.sartorius-mechatronics.com [Downloads].

4.4.8.1

Status Display

Requirements

In

-[Display items] the items are configured, see Chapter 5.6.7.

In

-[HW-Slots] PR 1721/46 is selected.

Display

1

2

3

4

LED ---

--- ---

---

Advertising