PR 5230 Instrument Manual
Installing the Instrument and Plug-in Cards
Sartorius
EN-65
4.3.4
Connecting up 2 to 8 Load Cells using 6-Wire Connecting Cable
Connections are made via junction box PR 6130/.. and connecting cable PR 6135 or PR 6136.
Caution!
The cable colors shown above are applicable to the 4-wire PR 62.. series load cells and
to the connecting cable PR 6135/36.
Before connecting, check the assignment of cable colors in the load cell manual.
Recommendation
-
Install cable in steel pipe connected to earth potential.
-
The distance between the measuring cables and the power cables should be at least 1 m.
Load cell supply circuit
-
Load resistance of load cell circuit >75 ¥, e.g. 8 load cells of 650 ¥ each
-
Supply voltage is 12 V DC, for further data, see Chapter 16.5.