2 leds on the fielbus card, Leds, Legend – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 87: 3 connection
PR 5230 Instrument Manual
Installing the Instrument and Plug-in Cards
Sartorius
EN-87
4.4.8.2
LEDs on the fielbus card
Watchdog LED
Meaning
Frequency
Color
Flashing
1 Hz green
Module is initialized and ready for operation.
Flashing
2 Hz green
Module is not initialized.
Flashing
1 Hz red
Check error in ASIC and FLASH ROM, module is defect.
LEDs
1 2
3 4
LED 1
LED 2
LED 3
LED 4
Lit
- - -:
No link or power off.
Lit
- - -:
Offline, no connection.
Lit
- - -:
No function.
Lit
- - -:
No power or not initialized.
Lit
grn:
Link established.
Lit
grn:
Link established and
online connected.
Lit
grn:
Initialized, no error.
Flashing 1 Hz
grn:
Receiving/Transmitting data.
Flashing 1 Hz
grn:
Online, no connection.
Flashing 1 Hz
grn:
Diagnostic data available.
Flashing 2 Hz
grn:
Engineering tool to identify
is active.
Flashing 1 Hz
red:
Configuration error.
Flashing 3 Hz
red:
No station name,
no IP address
Flashing 4 Hz
red:
Internal error.
Legend
---
off
red
red
grn
green
4.4.8.3
Connection
• Take the cable (e.g. patch cable CAT5) into the instrument through
the metal sleeve of the cable gland, strip the insulation and mount
the delivered RJ-45 plug (see mounting informations).
• Put the RJ-45 plug into the RJ-45 connecting socket of the fieldbus
card.