Commissioning
PR 5230 Instrument Manual
EN-110
Sartorius
5.4.2
Selecting the Calibration Mode
You can choose between [New] and [Modify]:
New
Modify
Param
5.4.2.1
New Calibration
Open the menu via
-[Weighingpoint]-[Calib].
When you press [New], the data is set to default first and calibration is started.
You are prompted to confirm:
Press [Continue] for the default settings, or [Cancel] to cancel the
selection.
Default settings with [New]:
Weighingpoint/WP A/
Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.000000
Max at
1.000000
Not calibrated
Sensitivity
833.33
4.000000
CalcTest