Pr 5230 instrument manual commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 165

Advertising
background image

PR 5230 Instrument Manual

Commissioning

Sartorius

EN-165

5.13.2

Loading Configuration and Calibration Data into the Device

Caution!
All data which can be adjusted in the setup menu are overwritten!
If the file is loaded into several devices, changing the network settings and the host name

is indispensable!


Procedure:

1.

Open the CAL-switch in the device.

2.

Click on ‚Backup of Earom’ to open the

menu ‘Backup-/Restore’.

Following window appears:


3.

Click on ‚Durchsuchen’ (depending on Internet browser).

Advertising