10 technical data, 1 general data, 2 power connection 230 v ac – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 36: 3 power connection 24 v dc, Technical data, Options, Pr 5230 instrument manual
Options
PR 5230 Instrument Manual
EN-36
Sartorius
3.2.10
Technical Data
Note:
Further technical data of PR 5230 see Chapter 15.
3.2.10.1
General Data
Accuracy
<10000 e (Klasse III) according to OIML R76/EN 45501
Connection for the non Ex area via 4mm
2
screw terminals
Connection for the Ex area
via 4mm
2
screw terminals
Intrinsically safe output
Protection class Ex ib IIC
Potential equalization
Via screw terminal PE, see Chapter 3.2.8.4
3.2.10.2
Power Connection 230 V AC
Supply voltage
100…230 V AC ±10 %
Power consumption
16 VA
Frequency
50…60 Hz ±2 Hz
3.2.10.3
Power Connection 24 V DC
Supply voltage
24 V DC ±10 %
Power consumption
12 W
3.2.10.4
Output Parameter (intrinsically safe) for the Connection of Load Cells/Junction
Box/Weighing Platforms
The connection will be done via 7-pin plug (marked with ‚X100’ and ‚Exib’) on the weighing electronics board.
Uo
21.0 V between the lines
lo
143 mA
Po
1.55 W
Co
188 nF between the lines
Lo
0.12 mH