5 determining the dead load, Determining the dead load, Pr 5230 instrument manual commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 115
PR 5230 Instrument Manual
Commissioning
Sartorius
EN-115
5.4.5
Determining the Dead Load
Note:
If the liniarization is active (see Chapter 5.4.10):
After selection of the line ‚Dead load at’ the following tip is displayed:
Can not be changed here
while linearization is active
Weighingpoint/WP A/
Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Not calibrated
Sensitivity
833.33
4.000000
by load
by mV/V
CalcTest
To use the empty scale/hopper as dead load (normal case):
• discharge the scale/hopper.
• press [by load].
After
or confirmation of the change is displayed:
Set deadload
If the mV/V value of the dead load was calculated, or if it is known from the previous calibration, the value can
be overwritten by pressing [by mV/V].
Weighingpoint/WP A/
Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
0.00 kg
0.000000
Max at
3000.00 kg
1.000000
Calibrated at
3000.00 kg
1.000000
Sensitivity
833.33
4.000000
by load
by mV/V
CalcTest