PR 5230 Instrument Manual

Installing the Instrument and Plug-in Cards

Sartorius

EN-79

4.4.5.3

9-Pole D-Sub Connector ‚IN’

E.g. Phoenix Contact IBS RTC-T
Pin allocation acc. to

DIN 41642

Signal

Color

DIN 47100

Description

Cable sheath

green

special InterBus cable (certified)

1 -----------------------------

IDO1

green

inverted data output

2 -----------------------------

DO1

yellow

not inverted data output

3 -----------------------------

IDI1

pink

inverted data input

4 -----------------------------

DI1

grey

not inverted data input

5 -----------------------------

GND

brown

signal - ground

6 -----------------------------

S

Screen

4.4.5.4

9-Pole D-Sub Connector ‚OUT’

E.g. Phoenix Contact IBS RTC-T
Pin allocation acc. to

DIN 41642

Signal

Color

DIN 47100

Description

Cable sheath

green

special InterBus cable (certified)

7 -----------------------------

IDO2

green

inverted data output

8 -----------------------------

DO2

yellow

not inverted data output

9 -----------------------------

IDI2

pink

inverted data input

10 ----------------------------

DI2

grey

not inverted data input

11 --------- * only if necessary

GND

brown

signal - ground
(continuation jumper: 11-12)

12 --------- * only if necessary

C

Con_Test (continuation jumper: 11-12)

13 ----------------------------- S

Screen

4.4.5.5

Connecting Diagram for a Master with three Slaves

Caution!
With Bus OUT 11 and 12 must be bridged, if another slave follows. Thus the link is not

fitted at the last slave, see next page.

Caution!
The cable screens must be taken into the instrument through the cable glands and

connected to terminals 6 or 13 (see next page).
Before, during and after installation, make sure that the sealing rings are seated

correctly.