Commissioning

PR 5230 Instrument Manual

EN-138

Sartorius

5.5.10

xBPI Calibration with Built-in Weight

Prerequisites:
-

The xBPI protocol has been selected (see Chapter 5.5.1).

-

The weighing point ‚xBPI-scale’ has been selected (see Chapter 5.5.4).

-

The platform configuration has been executed (see Chapter 5.5.3).

-

The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:

[Configuration]-[Confirming adjust.]: [manual] (see Chapter 5.5.5).

-

The communication between instrument and platform is active.

Weighingpoint/

xBPI-Scale Setup

Calibration

Open with

.

Configuration

Select group of specification

Show device info

Weighingpoint/

xBPI-Scale Setup

Adjust with default weight

Adjust with intern weight

Select with

and enter with

.

Linearity

Default

User

The procedure is shown e.g. with the following message:

Weighingpoint/

xBPI-Scale Setup

Calibration status

Motor in motion The calibration progress is displayed.

Accept

ResError

Abort

Weighingpoint/

xBPI-Scale Setup

Calibration status

Difference display

Cal-Delta

0.0 g

--

1212.73 g

Select [Accept].

Accept

ResError

Abort

The data are saved and the instrument generates a corresponding message:

Weighingpoint/

xBPI-Scale Setup

Calibration status

complete

Net=Grs

-0.0 g

--

-0.06 g

Leave with

.

Accept

ResError

Abort