Commissioning

PR 5230 Instrument Manual

EN-152

Sartorius

2. Determining a condition
Additionally, a condition [Condition] can be assigned to the marker.
Selection list for conditions [condition]

Condition

SPM bit

Description

no condition

-----

No condition

actual diginp1

X00=0

Digital input 1: not active

actual diginp2

X01=0

Digital input 2: not active

actual diginp3

X02=0

Digital input 3: not active

actual limit 1

X16=0

Limit signal 1: not active

actual limit 2

X17=0

Limit signal 2: not active

actual limit 3

X18=0

Limit signal 3: not active

ADC error

X32=0

General error in the weighing point: not active (no error)

above Max

X33=0

Weight above Max: not active

overload

X34=0

Weight above Max plus the ‘overload’ value: not active

below zero

X35=0

Weight not below zero

center zero

X36=0

Weight not within ¼ d of zero

inside ZSR

X37=0

Weight not within zero-setting range

standstill

X38=0

No mechanical stability of the scale

out

X39=0

Weight not below zero or above Max

command error

X48=0

For internal use only.

command busy

X49=0

For internal use only.

power fail

X50=0

Set after power-on (=power failure): not active

test active

X56=0

Analog test was not started.

cal active

X57=0

For internal use only.

tare active

X58=0

Instrument is not tared.

marker bit 1

X64=0

Marker bit 1 not set, after power-on the markers are set to ‘0’.

marker bit 2

X65=0

Marker bit 2 not set, after power-on the markers are set to ‘0’.

marker bit 3

X66=0

Marker bit 3 not set, after power-on the markers are set to ‘0’.