Commissioning

PR 5230 Instrument Manual

EN-146

Sartorius

[-none-]
Nothing is printed. The function can be used for printing less than 6 data items.

[Displayed weight], [Grossweight], [Net weight], [Tare weight]
The displayed gross, net or tare weight is printed. If [OIML], [NTEP] or [NSC] has been selected, printing is done
only, if the stability criteria is fulfilled, the weight is shown in '< >'. For [NTEP] or [NSC] the gross weight is
indicated with G (else B).

[Date

& Time]

The date and time are printed as DD.MM.YYYY HH:MM:SS
Example for printing

PrintlayoutItem1

Date

& Time

PrintlayoutItem2

Sequencenumber

PrintlayoutItem3

Grossweight

PrintlayoutItem4

Net weight

PrintlayoutItem5

CR/LF

PrintlayoutItem6

-none-

5.6.5

Fieldbus parameter

This menu item can only be selected if a fieldbus card is installed in slot 2.
Which protocol is displayed automatically depends on the Fieldbus card installed:
[ProfiBus-DP] for PR 1721/41, [InterBus-S] for PR 1721/42, [DeviceNet] for PR 1721/44, [ProfiNet I/O] for PR
1721/46 and [EtherNet-IP] for PR 1721/47.

Define the fieldbus parameters under

-[

Fieldbus parameter

].

Which additional parameters are required, depends on the interface type.
Press

to select the specific parameter.

Press

to select the respective values.

[ProfiBus-DP]

Setup/

Fieldbus parameter

fieldbus protocol

ProfiBus-DP

ProfiBus-DP address

1 Enter address

[InterBus-S]

Setup/

Fieldbus parameter

Fieldbus protocol

InterBus-S

[DeviceNet]

Setup/

Fieldbus parameter

Fieldbus protocol

DeviceNet

DeviceNet baudrate

500k Select 500, 250 or 125 k

DeviceNet MAC-ID

1 Enter address 1 ... 62