Commissioning

PR 5230 Instrument Manual

EN-106

Sartorius

5.3.8.1

Weighingpoint ‚Internal A’

- Calibration

Calibration of weighing electronics

- New

Reset Span and dead load

Contin, Cancel

- Max

0.00001 ...<3000>... 999999 <kg>, t, lb, g

- Scale interval

<1>, 2, 5, 10, 20, 50

- Dead load at

<0.000000 mV/V> or [by load]

- Max at

<1.000000mV/V> or [by load]

by load

0.00001 ... 999999 <kg>, t, lb, g

- Calibrated at

(Display only)

- Sensitivity (µV/d)

(Display only)

- Test

Determine test value

Exit calibration

Save or cancel changes

- Modify

Only for minor modifications/ setting new dead load, otherwise [New]

see New

- Param

- Measuretime

5 ms, 10, 20, 40, 80, 160, <320>, 640, 960, 1280, 1600ms

- Digital filter

<off>, Bessel, aperiod., butterw., tscheby.

- External supply

<8 -12 VDC>, below 8 VDC

- * Fcut

Cut off frequency, *only unless filter 'off', 0.1 - 80.0 Hz

- Test mode

<Absolute>, relative

- W

& M

<none>, OIML (cannot be selected, if 'Range mode: Multi-interval' has

been selected or Max has more than 3 decimals), NSC, NTEP

- Standstill time

0.01 s…<0.50 s> ... 2.0 s (range is dependent on response time)

- Standstill range

0.00 d ... <1.00 d> ... 10.00 d

- Tare timeout

0.1 s ... <2.5 s> ... 25 s, timeout due to instability

- Zeroset range

0.00 d ... <50.00 d> ... 10000.00 d

- Zerotrack range

0.00 d ... <0.25 d> ... 10000.00 d

- Zerotrack step

0.00 d ... <0.25 d> ... 10.00 d

- Zerotrack time

<0.0 s> ... 25 s

- Overload

0 d ... <9 d> ... 999999 d

- Min

0 d ... <50 d> ... 999999 d, minimum weight

- Range mode

<Single range>, multiple range, multi-interval (not if OIML selected)

- * Range limit 1

In weight, unit same as Max, transition from small to medium scale

interval

- * Range limit 2

In weight, unit same as Max, transition from medium to high scale

interval

* only for multiple range or multi-interval

- View(when CAL switch closed)

- Max

(Display only)

- Scale interval

(Display only)

- Dead load at

(Display only)

- Max at

(Display only)

- Calibrated at

(Display only)

- Sensitivity (µV/d)

(Display only)

- Param

Items as for Param. (display only)