J-Bus/ModBus Protocol

PR 5230 Instrument Manual

EN-168

Sartorius

6.2

ModBus-RTU

A serial interface is used for connection.
The telegram consists of four blocks:

Addr

8

Func

8

<data>

Crc

16

Legend

Addr

8

Slave device address within 1…126

Func

8

Function code

<data>

For further data, see Chapter 6.4

Crc

16

Checksum of all preceding characters


Note:

A telegram sent to Slave 0 is executed by all ModBus participants, but replied by none of these
devices!


At 9600 bauds, the reply time is typical 4 ms and maximal 8 ms.
A faulty command received by PR 5230 (e.g., parity error in the data, or CRC error) is ignored and no reply is
sent.
The pauses between the individual characters and a command must not exceed the 3.5-fold value of a
character length; otherwise PR 5230 detects a premature end of the command.