5 communication error, Communication error, Pr 5230 instrument manual sma protocol – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 185
PR 5230 Instrument Manual
SMA Protocol
Sartorius
EN-185
Example: 6000 kg x 1 kg platform scale
Command
Reply
<LF> I <CR>
<LF>SMA:2/1.0 <CR>
<LF>N<CR>
<LF>TYP:S <CR>
<LF>N<CR>
<LF>CAP:kg_:6000:1:0 <CR>
<LF>N<CR>
<LF>CMD:HPTMCR <CR>
<LF>N<CR>
<LF>END: <CR>
Example: 5000g x 1 g, 10000g x 2, 25000 g x 5 multiple range/multi-interval
Command
Reply
<LF> I <CR>
<LF>SMA:2/1.0 <CR>
<LF>N<CR>
<LF>TYP:S <CR>
<LF>N<CR>
<LF>CAP:g__:5000:1:0 <CR>
<LF>CAP:g__:10000:2:0 <CR>
<LF>CAP:g__:25000:5:0 <CR>
<LF>N<CR>
<LF>CMD:HPTMCRQ <CR>
<LF>N<CR>
<LF>END: <CR>
7.5
Communication Error
When a communication error due to a parity error or a frame error (if used) was detected, the scale sends an
ASCII ‘!’ character. The only other error is the detection of an unknown or non-supported command. In this
case, an ASCII '?' character is sent as a reply. Dependent on the error messages, the controlling computer must