SMA Protocol

PR 5230 Instrument Manual

EN-178

Sartorius

Requesting the High-Resolution Weight with Stability

Command: <LF>Q<CR>
Reply:

The scale returns the weight and status only, when the stability condition is met: Gross weight if
not tared, net weight if tared.

The stability condition must be met for this function. The maximum waiting time for stability is
set under [Tare timeout]; see Chapter 5.4.13.

<LF><s><r><n><m><f><xxxxxx.xxx><uuu><CR>

After elapse of the waiting time without reaching stability, the following reply is sent:

<LF><_><1><n><_><f><----------><___><CR>

For details, see page 180.

Requesting the High-Resolution Weight Continuously
This is a command which generates non-requested replies, because it does not function according to the strict
command-and-reply pattern. After the command, the scale repeats the reply continuously.

Command: <LF>S<CR>
Reply:

The scale repeats the high-resolution weight and status information continuously, until another
command is received.

<LF><s><r><n><m><f><xxxxxx.xxx><uuu><CR>

For details, see page 180.

Dependent on the used baud rate, the repetition rate of reply telegrams is roughly as follows:
19200 Bd

100 ms

9600 Bd

110 ms

4800 Bd

170 ms

Requesting the Tare Weight

Command: <LF>M<CR>
Reply:

The scale returns the tare weight and signals the 'tared' status in the <n> status character.

<LF><s><r><T><m><f><xxxxxx.xxx><uuu><CR>

For details, see page 180.