16.4

Effect of Ambient Conditions

16.4.1

Environmental Conditions

Temperature range

Ambient temperature for operation

-10...+40 °C

Ambient temperature 'not legal for trade'

-10...+50 °C

Power-on temperature

0...+40 °C

Storage/transport

-20...+70 °C

Humidity

<95 %, no condensation, (acc. to IEC 68-2)

Protection type

IP 66 housing

Height

<2000 m

16.4.2

Electromagnetic Compatibility (EMC)

All data in compliance with NAMUR NE 21, EN 45501 and EN 61326.

Housing

Radio frequency electromagnetic fields
(80 – 3000 MHz)

EN 61000-4-3

10 V/m

Electrostatic discharge (ESD)

EN 61000-4-2

6/8 kV

Signal and control
lines

Fast transients (burst)

EN 61000-4-4

3 kV

Peak voltages (surge) 1.2/50 µs

EN 61000-4-5

2 kV

Conducted disturbances by radio frequency
(0.15 – 80 MHz)

EN 61000-4-6

10 V

Mains inputs

Fast transients (burst)

EN 61000-4-4

3 kV

Peak voltages (surge) 1.2/50 µs

EN 61000-4-5

1/2 kV

Conducted disturbances by radio frequency
(0.15 – 80 MHz)

EN 61000-4-6

10 V

Voltage dips

EN 61000-4-11

0/40/70 %
20/200/500 ms

Short interruptions

EN 61000-4-11

5 s

16.4.3

RF Interference Suppression

Electromagnetic emission

In acc. with EN 61326, limit value class A