Operations on the instrument, Extended functions, Pr 5230 instrument manual – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 208

Advertising
background image

Extended Functions

PR 5230 Instrument Manual

EN-208

Sartorius

12.2.2

Updating via a Point-to-Point Connection with DHCP Service

Function ‘Auto-IP’ is used.
Prerequisites:

-

Device and notebook/PC are connected to each other.

-

The automatic address assignment ‘DHCP’ is activated in the device and in the

notebook/PC, see Chapter 18.2/18.3.

-

The ‘FlashIt!32’ program (in a directory on the enclosed CD-ROM) is installed on the

notebook/PC.

-

The ‘FlashIt!32’ program is started.

Operations on the Instrument

Danger! High Voltage!
Working on the instrument while it is switched on may have life-threatening

consequences.
The operation may be performed only by trained and qualified persons who are familiar

with the implied hazards and avoid them, or take adequate measure to protect

themselves.

1

2

3

1.

Open the door of the instrument.

2.

Press keys (1) and (3) simultaneously and press key (2) three times.

On the display appears:

3.

Press the key (1) to confirm the loading of the software.

On the display appears:

Now the device is ready to load the software.