Extended Functions
PR 5230 Instrument Manual
EN-208
Sartorius
12.2.2
Updating via a Point-to-Point Connection with DHCP Service
Function ‘Auto-IP’ is used.
Prerequisites:
-
Device and notebook/PC are connected to each other.
-
The automatic address assignment ‘DHCP’ is activated in the device and in the
notebook/PC, see Chapter 18.2/18.3.
-
The ‘FlashIt!32’ program (in a directory on the enclosed CD-ROM) is installed on the
notebook/PC.
-
The ‘FlashIt!32’ program is started.
Operations on the Instrument
Danger! High Voltage!
Working on the instrument while it is switched on may have life-threatening
consequences.
The operation may be performed only by trained and qualified persons who are familiar
with the implied hazards and avoid them, or take adequate measure to protect
themselves.
1
2
3
1.
Open the door of the instrument.
2.
Press keys (1) and (3) simultaneously and press key (2) three times.
On the display appears:
3.
Press the key (1) to confirm the loading of the software.
On the display appears:
Now the device is ready to load the software.