Commissioning

PR 5230 Instrument Manual

EN-154

Sartorius

5.8

Digital Outputs and Inputs

5.8.1

Configuring Digital Outputs

Configure the required function for [Output 1] to [Output 3] by selecting a signal from the list. The output goes
to the corresponding state, see Example.

Press

and select [Digital I/O parameter] to open the configuration menu.

Setup/

Digital I/O parameter

Output 1

below zero

X35=1

Output 2

above Max

X33=0

Output 3

center zero

X36=1

Input 1 on

-no action-

[Output 1] is true (active), when the weight value drops below zero (X35=1).
[Output 2] remains (active), as long as the weight is not above Max (X33=0).
[Output 3] is true (active), when the weight is zero ±¼ d (X36=1).
Selection list for output functions

Function

SPM bit

Description

actual diginp1

X00=0

Digital input 1: not active

actual diginp2

X01=0

Digital input 2: not active

actual diginp3

X02=0

Digital input 3: not active

actual limit 1

X16=0

Limit signal 1: not active

actual limit 2

X17=0

Limit signal 2: not active

actual limit 3

X18=0

Limit signal 3: not active

ADC error

X32=0

General error in the weighing point: not active (no error)

above Max

X33=0

Weight above Max: not active

overload

X34=0

Weight above Max plus the ‘overload’ value: not active

below zero

X35=0

Weight not below zero

center zero

X36=0

Weight not within ¼ d of zero

inside ZSR

X37=0

Weight not within zero-setting range

standstill

X38=0

No mechanical stability of the scale

out

X39=0

Weight not below zero or above Max

command error

X48=0

For internal use only.

command busy

X49=0

For internal use only.

power fail

X50=0

Set after power-on (=power failure): not active

test active

X56=0

Analog test was not started.

cal active

X57=0

For internal use only.

tare active

X58=0

Instrument is not tared.

marker bit 1

X64=0

Marker bit 1 not set, after power-on the markers are set to ‘0’.

marker bit 2

X65=0

Marker bit 2 not set, after power-on the markers are set to ‘0’.

marker bit 3

X66=0

Marker bit 3 not set, after power-on the markers are set to ‘0’.