Display of limits and digital inputs/outputs, Vnc display, Instrument display – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 157: Pr 5230 instrument manual commissioning, 9 display of limits and digital inputs/outputs

Advertising
background image

PR 5230 Instrument Manual

Commissioning

Sartorius

EN-157

2. Determining a condition
The selected action of each digital input can be combined with a condition that must be met for signal change
from 0 to 1 (on) or for signal change from 1 to 0 (off). Select the condition from the list; see page 152. No
condition is defined when selecting [no condition]; the action is executed directly.
Example: Taring via the digital input only if the gross weight exceeds the limit value.

Press

and select [Digital I/O parameter] to open the configuration menu.

Setup/

Digital I/O parameter

Output 1

marker bit 1

X64=1

Output 2

marker bit 2

X65=1

Output 3

marker bit 3

X66=1

Input 1 on

set tare

X113=1

Condition

tare active

X58=0

Input 1 off

-no action-

Input 2 on

-no action-

Input 2 off

-no action-

Input 3 on

-no action-

Input 3 off

-no action-

BCD out

Gross

In this example: If input 1 changes from 0 to 1 [input 1 on], a taring signal is triggered only if the condition under
[Condition] is met (limit 1 out = active).

5.9

Display of Limits and Digital Inputs/Outputs

5.9.1

VNC Display

Info/

HW-Slots

Builtin

Digital i/o

In use by PLC task

2

Digital outputs

001 The status is indicated from right to left.

Digital inputs

000

5.9.2

Instrument Display





The status is indicated from left to right.