PR 5230 Instrument Manual

SMA Protocol

Sartorius

EN-179

Controlling the Scale

Request for Taring of the Scale

Command: <LF>T<CR>
Reply:

The scale makes a taring attempt and signals the tared status in the <s> and <n> characters.

For this function, the stability condition must be met. The maximum waiting time for stability is
set under [Tare timeout]; see Chapter 5.4.13.

<LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR>

For details, see page 180.

Request for Taring with Fixtare Value

Command: <LF>T<xxxxxx.xxx><CR>
Reply:

The scale makes a taring attempt using the fixtare value <xxxxxx.xxx> and signals the tared
status in the <s> and <n> characters

<LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR>

For details, see page 180.

Request for Zero Setting of the Scale

Command: <LF>Z<CR>
Reply:

The scale makes a zero setting attempt and signals the zero status in the <s> character.

For this function, the stability condition must be met. The maximum waiting time for stability is
set under [Tare timeout]; see Chapter 5.4.13.

<LF><Z><r><n><m><f><xxxxxx.xxx><uuu><CR>

For details, see page 180.

Unless the scale is in the zero setting range, an error reply is generated.

Request for Tare Resetting

Command: <LF>C<CR>
Reply:

The scale deletes the tare weight and signals the tare reset status in the <n> status character.
The scale tare is reset.

<LF><s><r><G><m><f><xxxxxx.xxx><uuu><CR>

For details, see page 180.

Scale Diagnosis

Command: <LF>D<CR>
Reply:

The scale starts the diagnosis and returns a diagnosis reply.

<LF><r><e><c><m><CR>

For details, see page 180.