9 pr 1721/47 ethernet-ip interface, Pr 1721/47 ethernet-ip interface, Installing the instrument and plug-in cards – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 88: Pr 5230 instrument manual

Advertising
background image

Installing the Instrument and Plug-in Cards

PR 5230 Instrument Manual

EN-88

Sartorius

4.4.9

PR 1721/47 EtherNet-IP Interface

The EtherNet/IP interface PR 1721/47 is a plug-in card for mounting in slot 2, see Chapter 4.4.1. The card is
fitted with a standard RJ

-

45 socket for Ethernet connection and contains a powerful TCP/IP and EtherNet/IP

connecting circuitry with transfer rates of 10 and 100 Mbits/s.

Internal connection

34- pin connector on flat cable for
slot 2

External connection

RJ-45 connecting socket

Transfer rate

10 Mbit/sec and 100 Mbit/sec
Autodetection (10/100, HalfDX/FullDX)

Connection mode

Network

Protocol

EtherNet-IP

Configuration

EDS file ‚sag_5230_ethernetip.eds’

Cable

Twisted pairs, screened
e.g. patch cable CAT5
Autolink (straight or crossover)

Cable impedance

150 ¥

Cable length to HUB

Max. 115 m

Potential isolation Yes

Certificate

EtherNet-IP Specification
ODVA File No. 10286
Test Date: 06.09.2005
Vendor ID 90
See also: www.odva.org
Tested according to: CE, UL & cUL

Dimens. (LxWxH) 87 x 55 x 15mm
Weight

125 g


Note:

The IP-Address and the Subnet-Mask are set at Fieldbus parameter (see also Chapter 5.6.5 and 9).
The EDS file is stored on the CD (directory ‘Fieldbus’ of the according instrument) supplied with
the unit. The current file is also available for download via the Internet:
http://www.sartorius-mechatronics.com [Downloads].