PR 5230 Instrument Manual

Commissioning

Sartorius

EN-117

5.4.6

Calibration with Weight [by load]

Select [by load] for calibration using weight.

The weight value for the calibration weight must be entered in
separate window.

After applying the weight, enter the weight value and confirm with

. The weight unit for the calibration

weight (press

to change) may differ from the unit in the instrument; conversion is automatic.

Afterward, the following message is displayed:

Setting SPAN by load

Weight value, weight unit and measuring signal in mV/V corresponding to this value are displayed in the
[Calibrated at] line.

Weighingpoint/WP A/

Calibration

Max

3000 d

3000 kg

Scale interval

3000 d

1 kg

Deadload at

165.11 kg

0.057920

Max at

3000.00 kg

1.052369

Calibrated at

2000 10kg

0.701579

Sensitivity

876.97

4.209600

by load

by mV/V

by data

Linear.

CalcTest

This message displays, if the scale is not stable.
Remedial action
• Check the mechanical function of the scale.
• Adapt the filter setting; reduce the resolution, if necessary.
• Adapt the stability conditions.

This message displays, if the weight on the scale is less than the dead load
after input of the weight value.

The next step is calculation of the test value with [CalcTest] (see Chapter 5.4.10), and calibration is completed
with

(see Chapter 5.4.12).