Commissioning

PR 5230 Instrument Manual

EN-120

Sartorius

5.4.9

Subsequent Dead Load Correction

If the hopper/platform weight changes by an amount that is higher than the zero-setting range; e.g., due to
dead load reduction, dead load increase or mechanical changes, the functions for automatic zero tracking and
manual zero setting no longer work. To view the range which is already utilized by zero tracking or zero
setting, in [Calibration] press the

key; this also activates 10-fold increased resolution of the weight value.

Press

again to return to the previous state:

Current zero set: 0.123 kg

If the entire zero-setting range is already utilized, you can still correct the dead load (CAL switch must be
open) without affecting other calibration data/parameters.

To do this, select

-[Weighingpoint]-[Calib]-[Modify] and determine the dead load with [Deadload at]-[by

load] (see Chapter 5.4.5).
Note:

If the liniarization is active (see Chapter 5.4.10):
After selection of the line ‚Dead load at’ the following tip is displayed:

Can not be changed here

while linearization is active

5.4.10

Linearization

After selecting

-[Weighingpoint]-[Calib]- [New]/[Modify] and after completing calibration, select the

linearization menu with softkey [Linear.]:

by load

by mV/V

by data

Linear.

CalcTest


When you press [Linear.] the menu shown below appears:

Weighingpoint/WP A/Calibration/

Linearization

Max at

3000.00 kg

1.000000

Add

by mV/V

by load

To add a new linearization point, press [Add], fill in the weight value, apply the weight and press [by load].
Then fill in the corresponding value in mV/V for the weight. After pressing [mV/V], the value can be entered
directly.
Up to 3 linearization points can be determined.
A linearization point can be added with [Add], removed with [Delete] and changed with [Change].

Weighingpoint/WP A/Calibration/

Linearization

1. Lin. point

750 kg

0.250010

2. Lin. point

1500 kg

0.500020

3. Lin. point

2250 kg

0.750040

Max at

3000.00 kg

1.000000

Add

Change

Delete

by mV/V

by load

A linearization point can be selected with

, changed with [Change] and deleted with [Delete].