6 analog output (option pr 5230/06), 7 digital inputs, 8 digital outputs – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 223: 1 standard, 2 option, 9 serial interfaces, 1 rs-232, 2 rs-485, Analog output (option pr 5230/06), Digital inputs
PR 5230 Instrument Manual
Specifications
Sartorius
EN-223
16.5.6
Analog Output (Option PR 5230/06)
Output
gross or net or fixed preset value
Range
0/4 mA…20 mA
galvanically isolated
Load
0…500 ¥
Resolution
1 µA
Linearity error
<0,05 %
Total error
<0,1 %
Combined error - temperature
<100 ppm/K
16.5.7
Digital Inputs
3 inputs galvanically isolated via optocoupler (passive mode, external supply). Can be switched to active in the
instrument, for connection of switches. No common potential in passive mode.
Input signal for 'Low'
0…5 V or open input
Input signal for 'High'
10…28 V
Input current for 'High'
<11 mA @ 24 V, <5 mA @12 V
16.5.8
Digital Outputs
16.5.8.1
Standard
3 outputs, relays
Switch-over contact
Contact rating max.
250 V AC
5,0 A
1250 VA
250 V DC
0,3 A
100 V DC
0,5 A
50 V DC
1,5 A
30 V DC
5,0 A
16.5.8.2
Option
3 Opto-decoupled outputs
Supply voltage
24 V +10 %
Output current
Max. 40 mA
Voltage drop
3,2 V @ 40 mA
16.5.9
Serial Interfaces
16.5.9.1
RS-232
Transfer rate
Max. 19200 bd
Not galvanically isolated
16.5.9.2
RS-485
Übertragungsrate
Max. 19200 bd
Galvanically isolated