Register 2: state of state-controlled action bits, Register 4: calibration information, error byte, Pr 5230 instrument manual fieldbus interface – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 197

Advertising
background image

PR 5230 Instrument Manual

Fieldbus Interface

Sartorius

EN-197

Register 2: State of State-Controlled Action Bits
Only reading is permitted; the signal state is shown.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Byte 0
Byte 1
Byte 2 87:

GetFixTare

86:
SetFixTare

85:
ResetPWF

84:
ResetTest

83:
SetTest

82:
ResetTare

81:
SetTare

80:
SetZero

Byte 3

89:
ResetError

88:
Print

Register 4: Calibration Information, Error Byte
Only reading is permitted.

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Byte 0 EXPO
Byte 1 UNIT
Byte 2 STEP
Byte 3 LASTERROR

EXPO

One byte for the position of the decimal point; content in decimal form 0 ... 255

0 = 000000

1 = 00000.0

2 = 0000.00

3 = 000.000

4 = 00.0000

5 = 0.00000

UNIT

One byte for the weight unit; content in decimal form: 0 ... 255

2 = g, grams 3 = kg, kilograms

4 = t, tons 5 = lb, pound

STEP

One byte for scale interval; content in decimal form: 0 ...255

1 = scale interval '1', 2 = scale interval '2', 5 = scale interval '5'

10 = scale interval '10', 20 = scale interval '20', 50 = scale interval '50'

LASTERROR

LastError byte; see also

CmdError

bit, number of

'last error'

:

31 = no stability was reached (e.g. when taring)

33 = negative weight value when taring and ’legal-for-trade’ mode on

47 = no zero setting; weight not within zero setting range

107 = no stability with Getfixtare