1 controls on fieldbus card, 2 status-leds on fieldbus card, 3 status display – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 74: Display, Legend, Installing the instrument and plug-in cards, Pr 5230 instrument manual
Installing the Instrument and Plug-in Cards
PR 5230 Instrument Manual
EN-74
Sartorius
4.4.4.1
Controls on Fieldbus Card
The terminating resistors can be switched on (ON) and off by pressing the
switch, see Chapter 4.4.4.4.
Caution!
The rotary switch settings will not use.
Make sure that the switches for mode address 1…99 are
set to position ‘0’!
The mode address will be set in the menu [Setup]-[Fieldbus Parameter].
4.4.4.2
Status-LEDs on Fieldbus Card
Watchdog LED
Meaning
Frequency
Color
Flashing
1 Hz green
Module is initialized and ready for operation.
Flashing
2 Hz green
Module is not initialized.
Flashing
1 Hz red
Check error in ASIC and FLASH ROM, module is defect.
4.4.4.3
Status Display
Requirements
In
-[Display items] the items are configured, see Chapter 5.6.7.
In
-[HW-Slots] PR 1721/41 is selected.
Display
1
2
3
4
LED ---
--- ---
---
1
2
3
4
Lit
- - -:
no function
Lit
grn:
Fieldbus is online, data
transmission is possible.
Flashing 1 Hz
red:
Input/output length
configuration error.
Lit
red:
Fieldbus is off-line, data
transmission is not possible.
Flashing 2 Hz
red:
User parameter error.
Flashing 4 Hz
red:
Error in ASIC
Legend
---
off
red
red
grn
green