Xbpi protocol [xbpi port, Asycom protocol [asycom, Pr 5230 instrument manual commissioning – Rice Lake PR5230 Fieldbus Transmitter User Manual

Page 143

Advertising
background image

PR 5230 Instrument Manual

Commissioning

Sartorius

EN-143

xBPI protocol [xBPI Port]

To configure the serial interfaces, press

-[Serial ports parameter]-[ xBPI-Port]

to select a RS-485 or RS-232 interface.
Press

to select the specific parameter.

Press

to select the s respective values.

The light gray displayed parameter cannot be changed.

Setup/

Serial ports

Printer

Slot1 RS232

Remote display

Builtin RS232

JBUS/MOD-Bus

-none-

SMA

-none-

xBPI-Port

Slot1..RS485

Click [Param].

Param

Only the baud rate and the stop bits are adjustable; the other parameters are fixed.

Setup/Serial ports/

Slot1 RS485

Assigned to

xBPI-Port

Baudrate

9600 bd 9600, 19200

Bits

8

Parity

odd

Stopbits

1 1, 2

Asycom protocol [Asycom]

To configure the serial interfaces, press

-[Serial ports parameter]-[Asycom] to select a RS-485 interface.

Press

to select the specific parameter.

Press

to select the respective values.

The light gray displayed parameter cannot be changed.

Setup/Serial ports/

Builtin RS485

Assigned to

Asycom

Protocol

Asycom V1 Select V1, V2 or V3*

Baudrate

9600 bd 300, 600, .... 19200

Bits

8 7, 8

Parity

even

Stopbits

1

Slave

A Select A - Z


* V1 = for old communication programs
V2 = for recipe controller
V3 = for OPC

The old EW command sets (e.g. PR 1612) for the Asycom protocol are described in Chapter 8.