1 legal-for-trade operation, Commissioning, Pr 5230 instrument manual – Rice Lake PR5230 Fieldbus Transmitter User Manual
Page 124
Commissioning
PR 5230 Instrument Manual
EN-124
Sartorius
5.4.13.1
Legal-for-Trade Operation
Under
-[Weighingpoint]- [Calib]-[Param]-[W
&M] you can choose between:
[none] a legal-for-trade mode [OIML], [NTEP] or [NSC].
[none]
[OIML]
[NTEP]
[NSC]
Gross weight display
B
B
G
G
Min. meas. signal
0,125 mV/V at
30000 d
0,125 mV/V at
3000 e
0,125 mV/V at
3000 e
0,125 mV/V at
3000 e
0,25 mV/V at
60000 d
0,25 mV/V at
6000 e
0,25 mV/V at
6000 e
0,25 mV/V at
6000 e
0,42 mV/V at
10000 e
0,42 mV/V at
10000 e
0,42 mV/V at
10000 e
If legal-for-trade operation is selected, the parameters (zero tracking etc.) must be set accordingly; they are
not checked.
The relevant CAL switches (see Chapter 5.1.1) must be sealed in the closed position.
Note:
If the multi-interval scale [Multi-interval] has been selected in the menu [Range mode], the
W&M mode [OIML] cannot be selected.
With [OIML] selection, the following message is displayed:
W&M mode OIML not
allowed as long as Range
mode is multi interval
When used in legal metrology, no more than 3 decimals are permissible.
If a max. value with more than 3 decimals has been entered during
calibration, the following message is displayed.
Moreover, the following message is shown, if [OIML] has been selected:
W&M mode OIML not
allowed as with max of
more than 3 decimals.
Note:
In W&M mode an invalid weight is shown without mass unit.