PR 5230 Instrument Manual

Commissioning

Sartorius

EN-141

Remote display protocol [Remote display]
Select the serial interface to which the remote display is connected and then select [Param] to define the
[Baudrate] and the remote display type [Mode] connected.
Press

to select the specific parameter.

Press

to select the respective values.

The light gray displayed parameter cannot be changed.

Setup/Serial ports/

Builtin RS485

Assigned to

Remote display

Protocol

Remote display

Baudrate

9600 bd 300, 600, 1200, 2400, 4800, 9600, 19200

Bits

7

Parity

even

Stopbits

1

Mode

multiple transmitters Several remote displays are connected.

Device ID

A Address of the instrument

Next Device ID

B Next address of the instrument

If only 1 instrument is connected to a remote display (normal case), [Mode] must be set to [single transmitter].
For PR 1577 remote display [PR 1577 mode] has to be set.
If more than 1 instrument is connected to 1 remote display, [Mode] must be set to [multiple transmitters].
At [Device Id] the own instrument address (A, B, C ...) has to be entered, at [Next Device Id] the address of the
subsequent instrument has to be entered

Press

to return to the menu ‚Serial ports’.