Fluke Biomedical Cufflink User Manual
Page 100

Cufflink
Operators Manual
3-26
C
u
ffPres
C
u
ffPeak
DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/
8
0
0
194
3.5
53.2
45.4
4.1
57.3
(90)
N
200
160
120
40
0
20
40
60
40 BPM
8
0
fcv054.eps
Returned Data
194, 3.5, 53.2, 45.4, 4.1, 57.3; varies, depending on
parameter values
Example
Simulate a neonatal arm with a blood pressure of 120/80 , a heart rate of 30 BPM, a gain of
50%, and a +12 mmHg shift.
Command
makearmneo bp=adamsneo-120/80,hr=160, pulseid--
=8,envgain=50,envshift--12
Results
The Analyzer draws the makearmneo graph and
begins arm simulation when the NIBP monitor starts
inflating the cuff and results are sent to the controller
when the cuff deflates.
C
u
ffPres
C
u
ffPeak
DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/
8
0
0
142
3.1
45.1
43.2
3.1
4
8
.1
(90)
N
200
160
120
40
0
20
40
60
30 BPM
G 50
S 12
8
0
fcv055.eps
Returned Data
142, 3.1, 45.1, 43.2, 3.1, 48.1; varies, depending on
parameter values