Simulating neonatal blood pressure – Fluke Biomedical Cufflink User Manual

Page 45

Advertising
background image

Operation

Simulating Neonatal Blood Pressure

2

2-17

Simulating Neonatal Blood Pressure

Neonatal blood pressure is simulated and the monitor is tested in the same way as for

Adult blood pressure, with appropriate cuff and value differences. Target values are also

chosen from the

Select BP

submenu.

Note

Use the neonate mandrel with a neonate cuff when simulating neonatal

blood pressure.

To simulate neonatal blood pressure:
1. From the

Select BP

submenu, highlight

ADAMS Neonate

and press the

Ent

key.

The

ADAMS Neonate Family Target Values

screen displays:

F1

F2

F3

F4

F5

*** ADAMS Neonate Family ***
Target Values
60/30 (40)

150/100 (115)

80/50 (62)

200/150 (165)

100/80 (75)

100/80 (75)

HtRate

AdjEnv

Print

ZeroPres

fcv103.eps

2. See Table 2-1 for desired neonate target blood pressure values.

3. Use the same steps as for Simulating Adult Blood Pressure:

Select the heart rate

Adjust the pressure envelope

Set zero pressure

Run the NIBP monitor testing sequence

4. Compare the monitor values with the target values displayed on the

Makearm

display.