Fluke Biomedical Cufflink User Manual

Page 118

Advertising
background image

Cufflink

Operators Manual

3-44

MedTester REC # 15

SEQUENCE: 4

OP CODE: TEO

DEVICE INFORMATION

PHYSICAL INSPECTION

LINE VOLTAGES

GROUND RESISTANCE:

.002 OHMS

LEAKAGE TESTS, EQUIPMENT PWR OFF

CASE EXT LEAD NORM POL CLSD GND

.0

µAMPS RMS

CASE EXT LEAD NORM POL OPEN GND

.0

µAMPS RMS

CASE EXT LEAD REV POL OPEN GND

.0

µAMPS RMS

LEAKAGE TESTS, EQUIPMENT PWR ON

EUT CURRENT DRAWN:

.5 amps

COMMENTS:

NEXT TEST DUE DATE:

USER TIME:

ELAPSED TEST TIME: 62 SECONDS

CASE EXT LEAD REV POL OPEN GND

.0

µAMPS RMS

CASE EXT LEAD NORM POL OPEN GND

.0

µAMPS RMS

CASE EXT LEAD NORM POL CLSD GND

.0

µAMPS RMS

TYPE: ADULT*NIBP

MODEL:

120.5

120.5 VOLTS RMS

.4

L1-L2

L1-GND

L2-GND

MANF:

SN:

LOC:DNI

CN: ADULT*NIBP

DATE: 6/05/93

TIME: 15:38:28

fcv063.eps

Figure 3-8. medBase Adult BP Test Results