Pressure testing, Leak testing – Fluke Biomedical Cufflink User Manual

Page 47

Advertising
background image

Operation

Pressure Testing

2

2-19

Note

The baseline heart rate for all arrhythmias is set to 80 BPM.

6. Compare the monitor values with the target values displayed on the

Makearm

display.

7. Press

F3 Print

to print the arrhythmia test results.

8. Press

F4 Draw

to display the pressure envelope pulse.

Pressure Testing

The analyzer enables the user to test for pressure leaks in the test system, measure general

pressure, and evaluate the monitor’s pressure release valve. All three of these functions

are accessed from the

Press

submenu:

F1

F2

F3

F4

F5

fcv099.eps

Leak Testing

The

Leak Test

option allows testing of NIBP equipment, including cuff and tubing, for

air leaks.

Note

If the NIBP device has an internal system leak test or one that vents the

cuff inflation pneumatic circuit to the atmosphere when idle, do not use

the Leak Test option. Rather, use the Manometer option to check for

internal system leaks. Refer to the NIBP monitor operators manual for the

recommended test protocol.

To test for pressure leaks in the NIBP test system:
1. From the

Press

submenu, highlight

Leak Test

and press the

Ent

key. The

Leak

Test

screen displays:

F1

F2

F3

F4

F5

*** Leak Test ***

Pressure actual

(mmHg):

-

0.75 Target

Pressure drop

(mmHg):

15.50 200

Elapsed time (min:sec): 1: 0
Leak rate (mmHg/min): 15.50

Reset

Utility

Start

Pump

Release

Pressure

fcv027.eps

From the

Leak Test

screen, several options can be selected.

2. If necessary, set the cuff pressure to zero, as described in Simulating Adult Blood

Pressure: Setting Zero Pressure, above.

Advertising